10/27/97
I am experiencing an excessive charging in my SEM. The tested samples
are conductive, the specimen mount, holder and SEM stage are all well
grounded, but charging persists, precluding succesful imaging. Secondary
and backscattered electron images are equally affected. These samples
do not exhibit charging when analyzed under the same conditions (AV) in
another SEM. Many different samples were tested in these two SEMs and
the results are consistent.
It appears that some instrumental factor is involved. I would
appreciate the input on the subject. Thank you very much in advance.
Chris Terlecki
Applied Analytical Sciences
ph: 714-434-6894
email: aas@pacbell.net
If indeed the specimen and stage are grounded, you should check that you
are not using excessive beam current. Also, a large condenser spot or a
huge final aperture can be the problem. Since BS too is affected grounding
of the scintillator cannot be the problem.
If a bare specimen mount does not charge, then the specimen itself is
likely the problem. The other SEM may use just a little less beam current
and a partially conducting specimen would then not charge up.
If the problem is specimen related, one of the more common problems would
be the umbrella effect;
where the specimen is well coated on the upper surfaces, but these surfaces
prevent a continuos good coating on the under sides. Visualise a stack of
cannon balls or a mushroom: Coating from above would not be effective.
Such specimens can be sputter coated by laying the pin type mount on the
side, coating and then turning the specimen on the opposite side of the
pine and giving it a second coating.
At least this is a more interesting problem then fiddling with digital
files. Oh, did you use that address "experts" deliberately? I understand
the word means 'squirt under pressure'.
Cheers
Jim Darley
ProSciTech Microscopy PLUS
PO Box 111, Thuringowa QLD 4817 Australia
Phone +61 77 740 370 Fax: +61 77 892 313
Great microscopy catalogue, 500 Links, MSDS, User Notes
************************ http://www.proscitech.com.au
----------
images. You first need to really determine if it is the specimen or SEM.
Same effect with entirely different specimens? If you move the same
specimen between the different SEMs, is it always only in the one SEM? Use
scan rotation - does the effect stay with the specimen or move with the
scan direction. I guess I'd suspect some fault in the scan generation if it
is really to do with the SEM - electrical fault in scan generation, or
fault in coils, or connection to coils?
Regards,
--
Larry Stoter
17, Rocks Park Road, Uckfield, E. Sussex, TN22 2AT, UK
email: LPS@teknesis.demon.co.uk
Phone/Fax: +44 (0)1825 767967
I presume that for some reason you cannot coat the sample even though
they are conductive. A wiff ie 5nm AuPd on the surface should overcome
the charging. Also, turn down the wick on your microscope ie low kV, low
beam current.
Patrick Echlin
Cambridge
pe13@cus.cam.ac.uk
capable of measuring the absrobed current, see if that is operating
normally. A disconnected wire to ground from the stage will cause any
sample to charge. Check continuity between the stage and the microscope
body. Move the stage around while checking; you may have an intermitten
contact.
-Scott Walck
walck@ppg.com
contamination. A small bit of insulating material at the final
lens or in the column can deflect (often erratically) the incident
beam.
Woody White, Electron Microscopist SEM/EDS/WDS
Work: Mcdermott Technology, Inc.
woody.n.white@mcdermott.com
http://www.mtiresearch.com/