Can anyone give me their formula or direct me to a reference on how to

accurately calculate the size of scale bar on photomicrographs?



Thanks,



Ruth

Ruth Yamawaki <yamawaki@leland.Stanford.EDU>


Easiest way to do this (for me, anyway) is as follows:



1. Think of a value for your scale bar, e.g. 1 micron (this should

always be a round figure, or reasonable fractions thereof),

2. Multiply this by the final print magnification of the micrograph,

3. This value is the length of the scale bar, in this case in microns.



Example:



If final print magnification is 22 300x,

and if you choose to have the scale bar representing 1 micron,

then scale bar length = 1 micron x 22 300 = 22 300 microns.

Then, as there are 1000 microns in a mm, the scale bar length on

the micrograph will have to be 22.3mm.



Clearly, depending on the magnification, sometimes you may

choose an impractical figure for the scale bar to represent.



If, for example you choose 1 micron for a magnification of 400x

then the scale bar would only be 400 microns (0.4mm) long. This is

crazy so how about choosing 20 microns? The scale bar would

then have to be 20 microns x 400 = 8000 microns = 8mm.



At the other end of the scale, if you have a magnification of

120 000x then a 1 micron scale bar would be 120mm - much too

long - so how about a 0.1 micron (or 100 nm) scale bar? Yes,

much better because then the scale bar would be 0.1 micron x

120 000 = 12 000 microns = 12mm.



PLEASE do not fall into the trap of doing it the other way around

and calculating what a bar of given length will represent. This

leaves you then with all your bars the same length (very neat, yes)

but representing crazy figures like 0.23 micron, 66 nm, etc!



Thanks for asking about this. Now, having written out the answer,

with examples, on the many occasions when I am asked the same

question by our users, I can simply pass on copies of this

message rather than do the explanation over and over again on the

board!



Good luck



Robin







Robin H Cross

Director : EM Unit, Rhodes University, Grahamstown, South Africa

R.Cross@ru.ac.za (or eurc@giraffe.ru.ac.za)

tel: +27 46 603 8168 - fax: +27 46 622 4377

http://www.ru.ac.za/emu/index.htm






Ruth



I have found it quite useful to print out an Excel spreadsheet table with

values on it. One axis can have magnifications in 1,000s (eg

1,2,3,4,5,6,7,8,9,10,20,30,40,50,60,70,80,90,100,200,300,400,500) and the

other axis can have fixed values to be represented (eg

10,20,50,100,500nm;1,2,5,10,20um). The table can then contain the values in

mms.



It's almost foolproof and easy to run off copies for students, especially

when they realise that you can add bits together on the magnification side

to give you a scale bar length for say 13k = 10k + 3k so just add the two

entries in the table to get the length for a 2um bar. I imported this into a

word document because I wanted to shade bits of the table to give

recommended lengths for scale bars but it works fine.



I may have the Word 6.0 for Windows PC document somewhere. If you want a

copy post an e-mail to me (see e-mail address at bottom)



Malcolm



Malcolm Haswell

Electron Microscopy

School of Sciences

Fleming Building

University of Sunderland

SUNDERLAND SR1 3SD

Tyne and Wear

UK



Tel (0191) 515 2872

e-mail: malcolm.haswell@sunderland.ac.uk