10/9/98
I am having a problem with Spurr resin. The
blocks are coming out very brittle. I am careful
when I weigh out the components and polymerize at
65C. What else would cause brittle blocks?
Another problem is that staining with 1% T blue
with 1% borax is very poor. Any suggestions would
be of help. Thanks.
--
Regards,
Gregory Rudomen
Technical Specialist
University Microscopy Imaging Center
State University of New York at Stony Brook
516-444-3126
Greg@umic.sunysb.edu
For example, I once allowed some specimen vials to remain open
overnight in a rotator - to enhance penetration. Bad idea.
Result: brittle blocks that shattered during trimming. Even traces
of moisture in Spurr's will cause this. In ordinary epoxies, this is
not a problem (Epon for example could be left open overnight
or several days without this problem developing).
John J. Bozzola, Ph.D., Director
Center for Electron Microscopy
Neckers Building, Room 146 - B Wing
Southern Illinois University
Carbondale, IL 62901
U.S.A.
Phone: 618-453-3730
Fax: 618-453-2665
Email: bozzola@siu.edu
Web: http://www.siu.edu/departments/shops/cem.html
uses), or one exposed to moisture (If you keep the bottles in a low
temperature for extended shelf life, let it warm before opening the bottle).
With bottles out of the box, I had no problem.
In my opinion, Spurr's has relatively poor cutting properties, even when the
block itself is not brittle. I usually successfully mix Spurr's (complete
homogeneous mixture including DMAE) with Epox 812 (complete homogeneous
mixture including DMP-30) to improve cutting properties, with minimal
sacrifice in Spurr's excellent viscosity. It's been very cooperative for
plant tissues.
Hello,
I apologize for providing incomplete information.
I usually mix them 1 to 1 by weight. But you can reduce the amount of Epox
812 mixture, if low viscosity is important, as in freeze-dried tissues. It's
been tested primarily on plant tissues, fixed chemically or lipophillized,
and worked well for both. It worked on some chemically fixed animal tissues
(small intestine & liver from a frog). I don't have much experience with
animal tissues and I am not sure about lipophillized animal tissues.
But, I can say generally Spurr's provides better infiltration, while Epox
812 offers better cutting properties (via better adherence to the tissue).
There has been no negative effect of mixing two, as far as I know. Thus, I
guess any tissue, which are not incompatible with Epox or Spurr's, can be
safely embedded in the mixture of the two.
-----------------------------------------------
I also forgot including the staining method with toluidine blue in the
previous message. If it is to stain 'thick section' for preliminary
observation, place a drop of your 1% toluidine onto the section, heat it on
a 70 C-hot plate until you see thin dry edge around the stain, wash the
stain gently with squirts of water.
For microautoradiographic purpose, I use following method. This applies to 1
um-sections affixed on a gelatin-coated slide.
1) Dip the slides in 10% paraformaldehyde for 1-2 min.
2) Rinse the slides in distilled water for 1 min (or rinse twice for 30 sec.
each).
3) Stain sections by dipping the slide for 1/2 to 1 min in 0.05%-0.1%
toluidine blue prepared in 0.2M Na-phosphate buffer (pH 7.0). (Stain may
precipitate over the time; shake and refilter it.)
4) Wash the slide in running water quietly for about 2 min or shorter period
(prolonged washing will result very faint staining).
5) Air-dry the slide in vertical position.
Seung-Geuk Shin
SUNY-ESF at Syracuse
sgshin@syr.edu
sgshin@syr.edu
relatively short shelf life (I usually replace mine after about 6 months or
a year). Also the anhydride hardener - NSA - is particularly prone to
absorbing moisture so if a stock bottle has been opened for a long period
and I have a problem I try an unopened one. In any event I always label
bottles with the date of arrival in the lab and the date of opening.
Some moisture may be driven out of the resin during polymerisation but not
if you seal the lids on polythene BEEM capsules.
How good is your temperature measurement? We have an old oven with a dial
setting that needs to be checked if anyone else has used it and often the
only way is to start polymerisation a couple of hours before I leave the lab
on a night, so that I can confirm the polymerisation temperature.
I hope this helps.
Malcolm Haswell
Electron Microscopy
School of Health Sciences
Fleming Building
University of Sunderland
SUNDERLAND SR1 3SD
malcolm.haswell@sunderland.ac.uk
Here is the recipe for Spurr's + Epox 812, I use.
Basically, it is 1:1 mixture (by weight) of Spurr's and Epox 812.
Each resin mixture must be completely homogenized, including accelerator,
before combining the two.
1. Prepare Spurr's in a 100 ml disposable beaker.
(Based on Spurr's original formula)
VCD -------- 10 g
DER 736 ---- 6 g
NSA -------- 26 g
DMAE ------- 0.4 g
-------------------
Total ------ 42.4 g
2. Prepare Epox 812 in a 50 ml disposable beaker.
(Based on Luft's original formular, except for reduced DMP-30)
Epox 812 (WPE = 155*) --------- 25 g
DDSA (MW = 266) -------------- 14 g
NMA (MW = 178) -------------- 11 g
DMP-30 ----------------------- 0.5 g
*If WPE is different, proportions of each component should be recalculated.
Calculations can be found at
http://www.emsdiasum.com/ems/techdata/68.html).
3. To make 1:1 mixture, add 42.4 g of Epox 812 mixture into the 100 ml
beaker containing Spurr's and mix throughly. Amount of Epox mixture can be
reduced for hard-to-infiltrate tissues.)
4. Infiltration for plant tissues.
Transitional solvent Resin Mixture Infiltration
(propylene oxide or Time (with
diethyl ehter) rotation or occasional swirring)
-------------------- -------------- ------------
3 part 1 part 2 hrs.
2 2 2 hrs.
1 3 4 hrs.
0 4 4 hrs.
0 4 6 hrs.
-----------------------------------------------------------------
5. Cure blocks at 60 C for 48 hrs.
Regards,
Seung-Geuk Shin
sgshin@mailbox.syr.edu